Home > Articles > All Issues > 2014 > Volume 3, No. 2, April 2014 >

Calibration of K Type Thermocouple for Measurement of Temperature in Laboratory Equipment

N J Shirbhate and G K Dhongade
Department of Mechanical Engineering, Government Polytechnic, Amravati, India.

Abstract— Thermocouples are widely use for measurement of temperature in laboratories and industries. The accuracy of measurement is affected by the procedure employed for calibrating the thermocouples. In this work laboratory equipment is developed for calibrating the thermocouples. A K type thermocouple is calibrated for the temperature range from 300 C to 1000 C and results are presented. It is observed that correction factor of 0.077 mV is needed for the results to match with the calibrating standards.

Index Terms— Thermocouples, K-Type, Calibration, Laboratory Equipment

Cite: N J Shirbhate and G K Dhongade, " Calibration of K Type Thermocouple for Measurement of Temperature in Laboratory Equipment," International Journal of Mechanical Engineering and Robotics Research, Vol. 3, No. 2, pp. 289-292, April 2014.